Abstract: [
Abstract ] Objective ToinvestigatethevalueofthechestX-rayanddigitalsubtraction
angiography ( DSA ) incentralvenousportaccesssystem (
CVPAS ) catheterdysfunction.Methods
Onehundredandforty-twocasesofcatheterdysfunctionofCVPASwerestudied.Weanalyzed
thecausesofcatheterdysfunctionbythechestX-rayandDSA , treatedthecausesandobserved
theeffect.Results Thecausesof98 ( 69.01% ) catheterdysfunctionpatientswerefoundbyDSA
andchestX-rayin142casesofcatheterdysfunction , including51casesoffibrinshellformation ,
13casesofcatheterbendinthebodies , 11casesofcatheterfracture , 8casesofPinch-offsign , 7
casesofthrombusformation , 5casesofabnormallocationofendcatheterand3casesofcatheter
fastenerloosenessorfallout.Wetreatedthe98casesofcatheterdysfunctionagainstthesecauses.
77patients'CVPAScouldcontinuedtobeusedand21patients'CVPASwereremoved.Conclusion
ThechestX-rayisthe mostsimpleandeffective methodtofindthecausesofcatheter
dysfunctionofCVPAS.ThemainreasonforCVPAScatheterdysfunctionisfibrinshellformation.
DSAandthechestX-raycaneffectivelyevaluatethecauseofCVPAScatheterdysfunction , which
isworthyofclinicalapplication.