不同内固定方式固定骶髂关节脱位的三维有限元分析

  

  1. 河北医科大学第三医院创伤急救中心,河北省骨科研究所,河北省骨科生物力学重点实验室,河北,石家庄 050051
  • 出版日期:2018-09-25 发布日期:2018-08-28
  • 作者简介:吴涛(1983-),男,河北正定人,河北医科大学第三医院副主任医师,医学博士,从事骨外科疾病诊治研究。
  • 基金资助:
    国家自然科学基金青年基金(81702139);河北省自然科学基金青年基金(H2016206291)

Comparison of different kinds of internal fixation for sacroiliac joint dislocation:a finite element analysis#br#

  1. Trauma Emergency Center of Third Hospital of Hebei Medical University, Orthopaedic Research
    Institution of Hebei Province, Key Laboratory of Orthopaedic Biomechanics of
    Hebei Province,  Shijiazhuang 050051, China
  • Online:2018-09-25 Published:2018-08-28

摘要: [摘要]〓
〖HTH〗目的〖HTSS〗〖KG*2〗应用三维有限元分析法比较3种不同内固定器械治疗骶髂关节脱位的生物力学稳定性。
〖HTH〗方法〖HTSS〗〖KG*2〗随机挑选1名健康成年男性,收集层厚1.0 mm骨盆CT图像,经过Mimics 10.01、Geomagic Studio 12、SolidWorks、Abaqus 6.111等软件处理,构建完整的骨盆模型及骶髂关节脱位模型,分别使用2枚骶髂螺钉、可调式微创接骨板、张力带接骨板固定骨盆后环,对骶骨加载500 N负荷,比较整个模型和内固定器械的整体位移值、应变值、应力分布以及脱位的缝隙分离值。
〖HTH〗结果〖HTSS〗〖KG*2〗上半身的重力经过骶骨翼、双侧骶髂关节、坐骨大切迹、髂骨弓状线到双侧髋臼。在500 N 垂直载荷下,固定后的模型的整体位移值依次是2枚骶髂螺钉1.377 mm、可调式微创接骨板为2.517 mm、张力带接骨板为5.132 mm,脱位缝隙的分离移位值依次是2枚骶髂螺钉0.887 mm、可调式微创接骨板为2.502 mm、张力带接骨板为5.117 mm。
〖HTH〗结论〖HTSS〗〖KG*2〗2枚骶髂螺钉固定骶髂关节脱位的稳定性优于可调式微创接骨板的稳定性,更优于张力带接骨板的稳定性。

关键词: 骶髂关节, 内固定器, 生物力学

Abstract: [Abstract] Objective〖HTSS〗〓To compare the biomechanical stability of three kinds of internal fixation for sacroiliac joint dislocation by threedimensional finite element analysis.
〖HTH〗〖WTHZ〗Methods〖HTSS〗〓A healthy adult male was randomly seclected and the CT images of pelvis(1.0 mm thickness) were collected. Intact pelvis model and sacroiliac dislocation model were made using software of Mimics 10.01, Geomagic Studio 12, SolidWorks, Abaqus 6.111. Two iliosacral (IS) screws, the minimally invasive adjustable plate(MIAP) and the tension band plate(TBP) were used to fix the posterior pelvic ring respectively, and a loading of 500 N was put on the sacrum. Total displacement, strain value, stress distribution and the shift of dislocation gap of the whole model and internal fixator were compared.
〖HTH〗〖WTHZ〗Results〖HTSS〗〓The stress of upper body passed through sacral wing, bilateral sacroiliac joint, the large sciatic notch, arcuate line of the iliac bone to the bilateral acetabulum. Under 500 N load, displacements of the fixed models were two IS screws model(1.377 mm), the MIAP model(2.517 mm), and the TBP model(5.132 mm). The shifts of dislocation gap were two IS screws model(0.887 mm), MIAP model(2.502 mm) and TBP model(5.117 mm).
〖HTH〗〖WTHZ〗Conclusion〖HTSS〗〓The stability of sacroiliac dislocation fixed with the MIAP was less than that fixed with the two IS screws, and better than that fixed with the TBP.

Key words: sacroiliac joint, internal fixators, biomechanics